ZAHNER® Solutions
CIMPS
Controlled Intensity Modulated Photo Spectroscopy
The Zahner CIMPS system based on LED arrays brings an additional advantage compared to the IMPS described in the literature: a control loop regulates light intensity and modulation and keeps it absolutely stable. The automatic comparison between set value and sensed intensity eliminates the influence of non-linearity, ageing and temperature drift. Furthermore, it allows the direct calibration of the light source in units of intensity (W/m2).
The CIMPS package consists of several components, working together in a plug & play application, including software and an overall calibration of the system. An ElectroChemical Workstation ECW IM6 or ZENNIUM acts as a Frequency Response Analyser and as a support unit (Potentiostat / Galvanostat) for the cell under test. The IM-series workstations are renowned for their high precision, ease of use and comprehensive software. A slave potentiostat, driven by the IM6/IM6e system, controls amplitude and modulation of the light source over a wide range of frequencies.
An optical bench carries the housing of the light source with the integrated photo-amplifier and the slide with the photo-electrochemical cell (PECC-1 and PECC-2). A photodiode sensor is mounted near the site of the cells light inlet, adjustable in height and angle. Depending on the application, you can chose from different types of light sources (see table).
The Zahner CIMPS system based on LED arrays brings an additional advantage compared to the IMPS described in the literature: a control loop regulates light intensity and modulation and keeps it absolutely stable. The automatic comparison between set value and sensed intensity eliminates the influence of non-linearity, ageing and temperature drift. Furthermore, it allows the direct calibration of the light source in units of intensity (W/m2).
The CIMPS package consists of several components, working together in a plug & play application, including software and an overall calibration of the system. An ElectroChemical Workstation ECW IM6 or ZENNIUM acts as a Frequency Response Analyser and as a support unit (Potentiostat / Galvanostat) for the cell under test. The IM-series workstations are renowned for their high precision, ease of use and comprehensive software. A slave potentiostat, driven by the IM6/IM6e system, controls amplitude and modulation of the light source over a wide range of frequencies.
An optical bench carries the housing of the light source with the integrated photo-amplifier and the slide with the photo-electrochemical cell (PECC-1 and PECC-2). A photodiode sensor is mounted near the site of the cells light inlet, adjustable in height and angle. Depending on the application, you can chose from different types of light sources (see table).
CIMPS Software OptionThe ZAHNER CIMPS measurement system has been designed to match the special demands of dynamic photo-electric investigations. Besides traditional electrochemical investigations supported by the Thales software package a variety of special photo electric measurements can be done easily with the CIMPS option:
- SC fill-factor efficency
- Maximum power
- Intensity modulated photocurrent spectroscopy (IMPS)
- Controlled IMPS (CIMPS)
- Intensity modulated photovoltage spectroscopy (IMVS)
- Controlled IMVS (CIMVS)
- Charge extraction after N. W. Duffy, L. M. Peter et. al.
- Light transient measurements
- Time domain measurements
- DC vs. intensity transfer functions
- Chopped light voltammetry (CLV manual)
More than IMPS...
Available CIMPS-Systems
| CIMPS-Systems | |
| CIMPS-1 | Complete CIMPS system with
|
| CIMPS-2 | Complete CIMPS system with
|
| CIMPS-3 | Complete CIMPS system with
|
| CIMPS-4 | Complete CIMPS system with
|
| CIMPS-abs | Absorption Option with
|
| CIMPS-emit | Emission Option with
|
| CIMPS-pcs | Photo Current Spectra Option with
|
| CIMPS-fit | Fast Intensity Transients Option with
|
| CIMPS-dtr | Dynamic Transmittance and Reflectance Option with
|
Main Differences between PP211 and XPot as Lightsource Supply
| PP211 | XPot | |
| Frequency range | 10 μHz ... 100 kHz | 10 μHz ... 250 kHz |
| Supply output range for LED light source |
±10 A, ±20 V | ±500mA, ±10 V |
Optional Hardware
| SEL033 | NIST-traceable calibration sensor For automatic calibration procedure at the customer's lab |
| SIMPECC | Demo-board Electronic board for simulation of a PECC cell loaded with a sample
SIMPECC manual |
| PECC-1 | PECC-2 | |
| Physical dimensions (W x D x H) | 60 x 25 x 80 mm | 60 x 25 x 80 mm |
| Optical window diameter | 20 mm | 18 mm |
| Optical window material | BK7 or Quartz | BK7 or Quartz |
| Working electrode diameter | max. 20 mm | max. 18 mm |
| Solid material | Teflon (PTFE) | Kel-F (PCTFE) |
| Reference electrode | Ag/AgCl | Ag/AgCl |
| Counter electrode | Pt coil | Pt coil |
| Gas inlet/outlet | No | Yes |
Available Light Sources for CIMPS
| Color | Dominant Wavelength (nm), typical | Spectral Halfwidth (nm), typical | Intensity (W/m2), typical | Part Code Number |
| White | (3200K) | - | 400 *) | WLR02 |
| White | (4300K) | - | 900 *) | WLC02 |
| UV | 365 | 15 | 35 | LW365 |
| UV | 375 | 10 | 60 | LW375 |
| UV | 385 | 10 | 55 | LW385 |
| UV | 400 | 9 | 192 | UVR02 |
| UV | 405 | 9 | 168 | LW405 |
| 430 | 39 | 200 | LW430 | |
| 435 | 30 | 170 | LW435 | |
| Royal Blue | 447 | 11 | 1000 *) | KBR02 |
| 450 | 13 | 140 | LW450 | |
| Blue | 470 | 17 | 600 *) | BLR02 |
| Cyan | 505 | 19 | 340 *) | CYR02 |
| 525 | 23 | 104 | LW525 | |
| Green | 530 | 20 | 340 *) | GRR02 |
| 565 | 25 | 2 | LW565 | |
| 570 | 9 | 20 *) | LW570 | |
| Amber | 590 | 10 | 80 *) | GER02 |
| Red Orange | 617 | 12 | 300 *) | ORR02 |
| 620 | LW620 | |||
| Red | 627 | 10 | 300 *) | RTR02 |
| 630 | 10 | 159 | LW630 | |
| 645 | 11 | 157 | LW645 | |
| 660 | 20 | 65 | LW660 | |
| 670 | 15 | 108 | LW670 | |
| 680 | 20 | 100 | LW680 | |
| 690 | 15 | 74 | LW690 | |
| 700 | 20 | LW700 | ||
| 735 | 16 | 238 *) | LW735 | |
| 740 | 18 | 210 *) | LW740 | |
| 750 | 30 | 220 *) | LW750 | |
| 760 | 30 | 240 *) | LW760 | |
| 780 | 40 | 210 *) | LW780 | |
| 790 | *) | LW790 | ||
| 810 | 21 | 270 *) | LW810 | |
| 830 | *) | LW830 | ||
| 850 | 25 | 420 *) | LW850 | |
| 870 | 440 *) | LW870 | ||
| IR | 890 | 32 | 165 *) | LW890 |
| 910 | 40 | 140 *) | LW910 | |
| 940 | 40 | 180 *) | LW940 | |
| 950 | *) | LW950 | ||
| 970 | 34 | 50 *) | LW970 | |
| 990 | 35 *) | LW990 | ||
| 1050 | 34 | 34 *) | LW1050 | |
| 1200 | 100 | *) | LW1200 | |
| 1300 | 100 | 93 *) | LW1300 | |
| 1450 | 100 | *) | LW1450 | |
| 1550 | 100 | *) | LW1550 | |
| Royal Blue to UV (switchable) | 370 428 470 |
12 12 16 |
10 35 50 |
BUVZ02 |
*) Max. output power only available with PP211 (CIMPS-2 and CIMPS-4)
Tuneable Light Sources
| TLS | TXLS | |
| Physical dimensions (W x D x H) |
190 x 150 x 134 mm | 295 x 165 (224*) x 143 mm *) with projecting parts |
| Light source | High power LEDs | High power LEDs |
| Monochromator | USB controller unit | USB controller unit |
| Wavelength | 430 - 730 nm | 365 - 1020 nm |
CIMPS Resources
Product Information
Download the latest information brochure about ZAHNER CIMPS system...
catalog.pdf (PDF)
Manual Add-on
Download the latest manual add-on about ZAHNER CIMPS system...
CIMPS.pdf (PDF)
Download the latest manual add-on about ZAHNER CIMPS-abs option for the examination of electro-chromic processes...
CIMPS-abs.pdf (PDF)
Download the latest manual add-on about ZAHNER CIMPS-emit option for emission measurements on electroluminescent devices...
CIMPS-emit.pdf (PDF)
Download the latest manual add-on about ZAHNER CIMPS-pcs option for the examination of solar cells, DSSC, OSC,...
CIMPS-pcs.pdf (PDF)
Download the latest manual add-on about ZAHNER CIMPS-fit option for fast intensity transients measurements...
CIMPS-fit.pdf (PDF)
Download the latest manual add-on about ZAHNER CIMPS-dtr option for dynamic transmittance / reflectance measurements...
CIMPS-dtr.pdf (PDF)
Product Information
Download the latest information brochure about ZAHNER CIMPS system...
catalog.pdf (PDF)Manual Add-on
Download the latest manual add-on about ZAHNER CIMPS system...
CIMPS.pdf (PDF)Download the latest manual add-on about ZAHNER CIMPS-abs option for the examination of electro-chromic processes...
CIMPS-abs.pdf (PDF)Download the latest manual add-on about ZAHNER CIMPS-emit option for emission measurements on electroluminescent devices...
CIMPS-emit.pdf (PDF)Download the latest manual add-on about ZAHNER CIMPS-pcs option for the examination of solar cells, DSSC, OSC,...
CIMPS-pcs.pdf (PDF)Download the latest manual add-on about ZAHNER CIMPS-fit option for fast intensity transients measurements...
CIMPS-fit.pdf (PDF)Download the latest manual add-on about ZAHNER CIMPS-dtr option for dynamic transmittance / reflectance measurements...
CIMPS-dtr.pdf (PDF)