Add-on modules for IM5/6/6e electrochemical workstations

Modern electrochemistry at high currents

in IMPEDANCE SPECTROSCOPY, CYCLIC VOLTAMMETRY, POLARISATION CURVES ...
for BATTERIES, FUEL CELLS, ELECTROLYSIS, ELECTRO PLATING ...


 
 
 
Power potentiostats and electronic loads today are indispensable tools in several fields of electrochemistry, for example in the research of fuel cells. The potentiostats of the EL series are designed to sink high currents up to 100A/25A at a total power dissipation of up to 300W/100W (EL300/EL101). Both modules may be connected in full cell or half cell schemes. Using an additional external power supply you easily can extend an electronic load of the EL series to a two-quadrant potentiostat, you need for example in the field of electrolysis.
The EL101/300 is controlled by an EPC40, a module which is plugged into an extension slot of the IM6/6e.  Each EPC40 can address up to 4 PP/EL modules. Up to 4 EPC40 cards can be installed in an IM6 system, 1 in an IM6e system. In total, an IM6 is able to control up to 16 EL101/300 devices, an IM6e up to four. Each potentiostat will hold all its values from one control access to the next one so that no potential or current disturbances can occur while scanning the potentiostats. If series measurements are performed with more than one EL module connected, spectra are taken from all modules in a definable order.
The EL series modules are implemented completely in the IM6/6e environment. Thus, all acquisition and analysis techniques that run on these systems can as well be applied with the EL modules. The installation of one or more EL devices will upgrade your IM6/6e to an even more versatile high-current electrochemical workstation.

 
Supported Methods
Software Module
Impedance spectroscopy
IM 
Simulation & model fitting
SIM 
Cyclic voltammetry
CV 
Polarisation curves
CV 
Multicell multitasking voltammetry
CV 
Arbitrary current/potential/time measurements
PVI 
Capacity/potential measurements
C/E 
Automatic series measurements
AS 
Electrochemical Noise
NS
SCRIPT
AS

 
Specifications
EL101
EL300
Operating modes
potentiostatic / galvanostatic
potentiostatic / galvanostatic
Potential range
±4V / ±12V
±4V / ±12V
Potential accuracy
0.25%
0.25%
Current range
0A ... 25A
0A ... 100A
Current accuracy
0.25%
1%
Power dissipation
100W @ Ta
100W @ Ta
300W water cooled
Frequency range
10mHz ... 10kHz
10mHz ...   3kHz @ 100A
10mHz ... 10kHz @   25A
Ambient temperature Ta
0°C ... 25°C
0°C ... 25°C
System requirements
IM5/6 + EPC40
IM5/6 + EPC40

For connection of an EL device an EPC40 module is required !



Web Site Navigation





For all kinds of questions feel free to contact us. We are looking forward to hearing from you, e.g., by  E-MAILMail
 

ZAHNER-elektrik GmbH & Co. KG
P.O.Box 1846 - D-96308 Kronach - Germany
Tel.: (+49)-9261-52005/6
Fax: (+49)-9261-51919
e-mail: support@zahner.de
Handelsregister Coburg A 2857
USt./VAT ID no.: DE 133 053 948

Last update: 20..11.2002 by HJS