The  brandNEW
Electrochemical Workstation
IM6eX
provides features not known in this class before. All this at a
remarkable low price.


 
 
Equipped with a Ultra Low Noise Potentiostat
a pair of
High CMRR Precision Amplifiers for current and potential, 
a flexible
Dual DDS FRA  and 
High-Resolution Differential ADCs 
the
IM6eX  is designed for all situations where current and/or potential with or without DC and/or AC potential are to be measured. Thus, the IM6eX  provides a highly sophisticated impedance spectrum analyzer but also measures time domain data such as temperature, pH, rotation speed and others. With 4 extension slots it is open for a great variety of hardware modules, being able to fit all applications optimally.

 
Main differences between IM6 and IM6eX
 
IM6eX
IM6
Upper frequency limit
3 MHz
8 MHz
Current range
+-2 A
+-3 A
Rise time

250 ns

100 ns
Extension slots 4
Integrated EPC40 no yes
PC interface USB USB

 
 Supported Methods

 

Method Module Option
Electrochemical Impedance Spectroscopy EIS -
EIS Series Measurements vs. Parameters EIS -
Impedance Network Analysis, Simulation & Fitting SIM -
Impedance/Parameter, Capacity/Voltage Curves C/E -
AC Voltammetry C/E -
Cyclic Voltammetry CV -
Steady State Current/Voltage Curves I/E -
Multi-Cell Current/Voltage Curves MI/E EPC40
Corrosion / Polarisation Measurement POL -
Chronoamperometry PVI -
Linear Sweep Voltammetry PVI -
Electrochemical Noise NOISE EPC40, NProbe
CorrElNoise NOISE EPC40, NProbe
Tast Polarography PGY -
Differential Pulse Polarography PGY -
Differential Pulse Voltammetry PGY -
Stripping Voltrammetry PGY -
Standard Addition Measurement & Analysis PGY -
Universal Measurement Data Acquisition & Control ACQ NET
Universal Frequency Response Analysis FRA -
AC-DC-AC Tests COLT COLT-Kit
Layer Quality Test / Bi-Layer Test COLT COLT-Kit
Fast Pulse & Transient Recording TRC TR8M
High Current Interrupt Measurements PULSE TR8M
Relaxation Voltammetry RV EPC40, NProbe
Controlled Intensity Modulated Photo-Spectroscopy CIMPS CIMPS
Programmable Procedures SCRIPT -

 

 S P E C I F I C A T I O N S  IM6eX
 Frequency Generator and Analyser  
 Frequency range  10 uHz to 3 MHz
 Resolution  5000 steps / decade
 AC-amplitude range  1 mV to 1 V
 Impedance range potentiostatic
                                     /w buffer
   1 mOhm -    1 GOhm   / +-2%
100 mOhm - 100 GOhm  /+-3%
Impedance range galvanostatic   30 uOhm  -    1 GOhm  /+-2%
 Input channel phase-tracking accuracy   +-0.1 deg @ 10 uHz to 100 KHz 
 +-0.25 deg @ 100 KHz to 3 MHz
 Common mode rejection (differential inputs)  > 80 dB at   10 uHz to 100 KHz
 > 60 dB at 100 KHz to    3 MHz 
 Internal Potentiostat    
 
DC Characteristics
 
 Counter electrode  compliance voltage  +-14 V  (up to +-120 V boosted)
   output current  +-2 A
 Continuous power    28 W max
 Potential ranges    +-1 V, +-2 V, +-4 V, +-10 V
 Resolution    64.000 steps / range
 Input impedance  without buffer  typical  1 TOhm  // +-5 pF
 min.    50 GOhm // +-5 pF
   with buffer  typical 10 TOhm // +-5 pF
 min.      1 TOhm // +-1 pF
 Input current  without buffer  10 pA
   with buffer  12 fA
     
 
 Dynamic Data
 
 Small signal rise time    250 ns to 200 us in 5 steps 
 automatic range selection
 Slew rate    15 MV/s
 Bandwidth  (-3 dB)     3 MHz
 Phase shift    10 deg at 200 KHz
     
 
 Potentiostatic Mode
 
 Potential range    +-  4 V nominal
 +-10 V extended
   resolution  125 uV (standard) / 320 uV (extended)
   accuracy  +-250 uV (standard) / +-1 mV (extended)
 Input current ranges    +-1 nA to +-2 A in 26 steps
 automatic range selection
   minimum resolution  0.1%
   accuracy  +-0.05% @ 2 uA to 200 mA
 +-0.5%  @ <2 uA or >200 mA
     
 
 Galvanostatic Mode
 
 Current range    +-0.1 uA to +-2 A nominal in 24 steps
   minimum resolution  0.025%
   accuracy  0.1% @  >2 uA to 200 mA
 1%  @ <2 uA or >200 mA



Web Site Navigation





For all kinds of questions feel free to contact us. We are looking forward to hearing from you.  E-MAILMail
 

ZAHNER- elektrik GmbH & Co. KG
Thüringer Str. 12  - 96317 Kronach - Germany
Tel.: (+49)-9261-962119-0
Fax: (+49)-9261-962119-99
e-mail: support@zahner.de
Handelsregister Coburg A 2857
USt./VAT ID no.: DE 133 053 948
Owners: C.A. Schiller, W. Schiller

Last update: 02.08.2007 by HJS