| Equipped
with a Ultra Low Noise Potentiostat, a pair of High CMRR Precision Amplifiers for current and potential, a flexible Dual DDS FRA and High-Resolution Differential ADCs the IM6eX is designed for all situations where current and/or potential with or without DC and/or AC potential are to be measured. Thus, the IM6eX provides a highly sophisticated impedance spectrum analyzer but also measures time domain data such as temperature, pH, rotation speed and others. With 4 extension slots it is open for a great variety of hardware modules, being able to fit all applications optimally. |
|
|
|
|
| Upper frequency limit |
|
|
| Current range |
|
|
| Rise time |
250 ns |
|
| Extension slots | 4 | 8 |
| Integrated EPC40 | no | yes |
| PC interface | USB | USB |
| Method | Module | Option |
| Electrochemical Impedance Spectroscopy | EIS | - |
| EIS Series Measurements vs. Parameters | EIS | - |
| Impedance Network Analysis, Simulation & Fitting | SIM | - |
| Impedance/Parameter, Capacity/Voltage Curves | C/E | - |
| AC Voltammetry | C/E | - |
| Cyclic Voltammetry | CV | - |
| Steady State Current/Voltage Curves | I/E | - |
| Multi-Cell Current/Voltage Curves | MI/E | EPC40 |
| Corrosion / Polarisation Measurement | POL | - |
| Chronoamperometry | PVI | - |
| Linear Sweep Voltammetry | PVI | - |
| Electrochemical Noise | NOISE | EPC40, NProbe |
| CorrElNoise | NOISE | EPC40, NProbe |
| Tast Polarography | PGY | - |
| Differential Pulse Polarography | PGY | - |
| Differential Pulse Voltammetry | PGY | - |
| Stripping Voltrammetry | PGY | - |
| Standard Addition Measurement & Analysis | PGY | - |
| Universal Measurement Data Acquisition & Control | ACQ | NET |
| Universal Frequency Response Analysis | FRA | - |
| AC-DC-AC Tests | COLT | COLT-Kit |
| Layer Quality Test / Bi-Layer Test | COLT | COLT-Kit |
| Fast Pulse & Transient Recording | TRC | TR8M |
| High Current Interrupt Measurements | PULSE | TR8M |
| Relaxation Voltammetry | RV | EPC40, NProbe |
| Controlled Intensity Modulated Photo-Spectroscopy | CIMPS | CIMPS |
| Programmable Procedures | SCRIPT | - |
| Frequency Generator and Analyser | |
| Frequency range | 10 uHz to 3 MHz |
| Resolution | 5000 steps / decade |
| AC-amplitude range | 1 mV to 1 V |
| Impedance
range potentiostatic /w buffer |
1 mOhm - 1 GOhm / +-2% 100 mOhm - 100 GOhm /+-3% |
| Impedance range galvanostatic | 30 uOhm - 1 GOhm /+-2% |
| Input channel phase-tracking accuracy | +-0.1
deg @ 10 uHz to 100 KHz
+-0.25 deg @ 100 KHz to 3 MHz |
| Common mode rejection (differential inputs) | >
80 dB at 10 uHz to 100 KHz
> 60 dB at 100 KHz to 3 MHz |
| Internal Potentiostat | ||
|
|
||
| Counter electrode | compliance voltage | +-14 V (up to +-120 V boosted) |
| output current | +-2 A | |
| Continuous power | 28 W max | |
| Potential ranges | +-1 V, +-2 V, +-4 V, +-10 V | |
| Resolution | 64.000 steps / range | |
| Input impedance | without buffer | typical
1 TOhm // +-5 pF
min. 50 GOhm // +-5 pF |
| with buffer | typical
10 TOhm // +-5 pF
min. 1 TOhm // +-1 pF |
|
| Input current | without buffer | 10 pA |
| with buffer | 12 fA | |
|
|
||
| Small signal rise time | 250
ns to 200 us in 5 steps
automatic range selection |
|
| Slew rate | 15 MV/s | |
| Bandwidth (-3 dB) | 3 MHz | |
| Phase shift | 10 deg at 200 KHz | |
|
|
||
| Potential range | +-
4 V nominal
+-10 V extended |
|
| resolution | 125 uV (standard) / 320 uV (extended) | |
| accuracy | +-250 uV (standard) / +-1 mV (extended) | |
| Input current ranges | +-1
nA to +-2 A in 26 steps
automatic range selection |
|
| minimum resolution | 0.1% | |
| accuracy | +-0.05%
@ 2 uA to 200 mA
+-0.5% @ <2 uA or >200 mA |
|
|
|
||
| Current range | +-0.1 uA to +-2 A nominal in 24 steps | |
| minimum resolution | 0.025% | |
| accuracy | 0.1%
@ >2 uA to 200 mA
1% @ <2 uA or >200 mA |
|